Application Note 04: Cropscan 2000G Wheat
Prediction
February 2000
Introduction
The Cropscan 2000G Whole Grain Analyser is provided with
calibrations designed to simultaneously measure protein and moisture content of
the major classes of wheat grown in Australia. An evaluation of the predictive
ability of these calibrations is presented here including statistical analyses.
NIR spectra of the different wheat classes are shown in the diagram below.
Description
26 Australian wheat samples were analysed in the laboratory for
protein (Leco) and moisture (Oven) content. These samples were scanned on the
Cropscan 2000G between 720-1100nm in a 20mm pathlength cell using the wheat
calibration provided in the instrument. Five individual scans were recorded for
each sample and the average was taken as the predicted protein and moisture
values.

Results
A graphical representation of the prediction results are shown
here


The following table shows the comparison of the results of the Cropscan 2000G against the laboratory data.

Conclusion
The above data shows that the Cropscan 2000G can be used to measure the
protein and moisture content of wheat samples to within standards specified by
the Australian Wheat Board. The Absolute Average Deviation (AAD) specification
for wheat are 0.33% for protein and 0.38% for wheat. The Cropscan 2000G
predicts these properties well within these limits (0.22% for protein and 0.17%
for moisture).