Technology Description
What is NIR?
Near Infrared (NIR) radiation lies in the spectral region between 750-2500nm in the
electromagnetic (EM) spectrum, ie. between the visible and mid-infrared (IR).

Parts of the electromagnetic spectrum used for spectroscopy.
In general, IR spectroscopy is concerned with the absorption of
radiation resulting in molecular stretching, bending and rotations.

Two Regions of the NIR
The NIR region of the EM spectrum is concerned with overtone and combination bands
associated with the mid-IR (fundamental) frequencies.

 | The Cropscan 2000G scans the range between 720-1100nm |
 | The NIT and FOP analyser scans between 500-1100nm |
Spectral Information
Understanding NIR spectral information serves 2 purposes:
- Allows prediction of where a particular chemical species should absorb.
- Provides an assessment of the ability of NIR to perform an application.
C-H Absorptions
The carboxy-group (CH) is fundamental to compounds such as
aliphatics and aromatics.
N-H Absorptions
The measurement of the N-H vibrations makes quantitative
measurement of proteins by NIR possible.
O-H Absorption
The hydroxy-group (OH) is fundamental to compounds such as
alcohols and water.
Fundamental Principles of Spectroscopy
Absorption of Light
The incident radiation Po passes through a layer of an
absorbing species of pathlength b and concentration c. As a consequence of
absorption, the power of the incident light beam is attenuated from Po to P.
Transmittance is defined by: 
Absorbance is defined by: 
Absorbance and Transmittance are both wavelength dependent quantities, therefore to
obtain a spectrum, absorbance must be measured at each particular wavelength the
instrument scans.
The Relationship Between Absorbance and Concentration
.
The Beer-Lambert Law
Where: A = absorbance
e = absorptivity
b
= pathlength
c
= concentration
Absorptivity e (sometimes referred to as the molar
extinction coefficient) related to the probability that a transition (either electronic or
vibrational) will occur.
The Beer-Lambert law is used extensively in the calibration and determination of
unknown species in solution using UV-Vis spectroscopy.
Modes of Sampling
Three modes of sampling will be discussed
- Transmission
- Diffuse Reflectance
- Interactance
Transmission
Refer to diagram for single beam instrument. The
Cropscan 2000G is a transmission instrument
 | 100% scan measured directly from lamp. |
 | Light beam passes through sample. |
 | Usually requires minimal sample preparation. |
 | Available as single or double beam configurations
|
Two types of transmission measurement should be considered:
 | Pure Transmission |
 | Transflectance |
Pure Transmission
This type applies to homogenous samples, such as liquids, where minimal scatter of the
incident radiation occurs and that measured at the detector is pure absorbance by the
sample.

 | Spectra obtained are usually reproducible and result in good
calibration data (depending on the accuracy of the reference method and whether the
property being measured can be related to its spectra) |
Transflectance
 | This type of scattering occurs when the sample packing cannot be
accurately reproduced. |
 | Sample must be scanned several times and an average of the packings it taken as the
final result |
 | Absorbance spectra are not purely sample characteristic absorptions, they also contain a
bulk absorption component related to sample packing density. |
 | Although sample absorbances may be as high as 4.5abs, the spectral base to peak height
is still usually within 1abs. |
 | The above principals apply to liquid samples containing microparticulate matter, which
can exaggerate scattering effect even more than whole grains. |
Modes of Sampling (cont.)
Diffuse Reflectance
This is the mode of operation used by the Fibre Optic Probe (FOP) Analyser.
Principle of Diffuse Reflectance


 | Requires reflectance standard to obtain 100% spectrum. |
 | Material must be homogenised or ground to a constant particle size for reproducible
measurements. |
 | US bureau of standards report that optimal light detected at 45°
to source. |
 | Reflected light is a combination of Specular reflectance Sp and Diffuse
reflectance RD. |
 | Must contain more diffuse reflectance information than specular to be useful. |
Modes of Sampling (cont.)
Interactance