Cropscan 1000B Whole Grain Analyser
![]()
![]()
Cropscan 1000B is Near
Infrared Transmission Analyser designed to measure protein and moisture in
whole grains of wheat and barley. The Cropscan 1000B also includes a Test Weight
measurement .
![]()
![]()
![]()
![]()
![]()
![]()

![]()
![]()
![]()
![]()
![]()
![]()
The Cropscan 1000B
Whole Grain Analyser uses a diode array
spectrometer to scan the wavelength region, 720-1100nm. Within this region of
the NIR spectrum, protein, moisture and starch absorb NIR energy. Also colour
can be measured by scanning the end of the visible spectrum. Light from a
tungsten halogen lamp is passed through a sample of grain. The light is
dispersed across a silicon photodiode diode array detector to produce a NIR
spectrum. The amount of light absorbed across the NIR spectrum is proportional
to the concentration of the protein and moisture.
![]()
![]()
![]()
Grain is poured into the specially made sample cup. A
blade is inserted across the grain and the excess grain is poured out. The
blade is removed from the cup and the 500ml of grain is poured into the sample
hopper. The grain is metered through the optical chamber where the NIR spectra
are collected and stored in memory. The average of between 15 and 30 spectra
are used to compute the percentage protein and moisture. The grain falls into
the sample cup which is located on a load cell. The weight of the 500ml of
grain is recorded and displayed along with the protein and moisture.
Features and Benefits
NIR
Transmission technology Same
NIR technology as used by AWB, Graincorp, CBH and Ausbulk
Flow
Through Sampling System Rapid
and simple to use
Optional
Test Weight Module Provides
Hectolitre Weight measurement
Solid State
Optics Unaffected
by vibration
Independent
of orientation
Rugged, stable
and compact
Four Button Keypad Simple
to use.
Built-in Autocalibration
Software Calibration
adjustments can be performed using a single test sample.
Internal
Computer, Keyboard, LCD Stores
calibrations and predicts constituents onto a LCD
Save results
using alpha/numeric characters
RS232
Serial Port Provides
a convenient method of uploading stored data
to a PC or
to download calibrations to the instrument
Small Footprint Requires
less bench space
Specifications
Scan Range: 720-1100nm
![]()
Constituents: Up to 6 constituents displayed
including, Protein, Moisture, Test Weight, Gluten, Sedimentation.
Pixels 38
Scan Speed: 2-4 seconds
Power: 110/240VAC,
18VDC
![]()
![]()
![]()
![]()
Weight and Dimensions: 12Kg, 330mm(W) x 300mm(D) x
400mm(H)
![]()
![]()
![]()
![]()
![]()
![]()