
Introduction:
NIR Transmission Spectroscopy provides a means of measuring C-H, O-H and
N-H in a broad array of plastics and polymers. This simple study was conducted to
demonstrate the ability of the NIT-38 Polymer Analyser to collect NIT spectra
of various plastic films, polymers and solvents.
Description:
Samples of Polyethylene, Polypropylene, Polystyrene, Polycarbonate films
were scanned using the NIT-38 Polymer Analyser. Samples of Xylene, Acetone and
Ethanol were scanned in a 20mm pathlength cuvette. A sample of Silicon Adhesive
was scanned as the material hardened on a glass slide.
Results:
Figure 1 through 4. shows the spectra of the plastics films.

Figure 1. Plot of NIT Spectrum of Polycarbonate Sheet, 5mm thickness.

Figure 2. NIT Spectrum of Polyethylene Bag, 4 pieces.

Figure 3. NIT Spectrum of Polystyrene Film, 4 pieces.

Figure 4. NIT Spectrum of Polypropylene Film, 4 pieces.

Figure 5. NIT Spectra of Solvents: Xylene(Red), Ethanol(Green Top),
Acetone(Green Lower)

Figure 6. NIT Spectra of Silicon Adhesive during hardening. The Red
spectrum is the final hardened material.
Comments:
This study is intended to demonstrate the spectral differences between
various plastics, polymers and solvents. Since the NIR spectra contain only
C-H, O-H and N-H absorbance bands, the spectra are relatively simple as
compared with Mid IR(FTIR) spectra. The advantage of using the NIR spectral
region is ease of sample presentation, ie, all these samples were scanned in
their natural state. Also the NIR spectra are easy to interpret and changes in
chemical bonds occurring during polymerisation reactions are easily detected
and quantified.
The spectral region 720-1100nm, as used in the NIT-38 Polymer Analyser,
is suitable for measuring thin films, ie, less than .5mm thick, and through
long pathlengths, ie, 20-30mm. The use of fibre optics in this region of the
spectrum is also an advantage as low cost fibre optic bundles can be used over
long distances.