Introduction:
.
Near Infrared
Spectroscopy, NIRS) measures fat, protein, sugars and water based on the
presence of C-H(fat), N-H(protein), C-O-H(sugars) and O-H(water) bonds in a
product such as ice cream. To
demonstrate the feasibility of NIRS to measure components in ice cream, 6
samples of ice cream and sorbet
products were purchased from the local supermarket. The study is not intended
to prove the method, but more to illustrate the potential for this application.
Description:
Samples of 6 different
brands and types of ice cream and desert products were selected to provide a
range of fat concentrations between 1 and 14%. Three samples were white and
three were chocolate flavoured. One of the chocolate samples was a sherbet, ie,
1% fat content. The samples were left at room temperature to bring them to a
liquid consistency.
100ml of each sample
was poured into a glass petri dish and loaded into the Series 3000 Food
Analyser. 10 Scans of each sample were collected between 720 and 1100nm. All
samples were scanned in duplicate. Figure 1. shows the NIT spectra of the
samples.
The spectra were
imported into Excel where the component levels were added to the spectral data.
The component levels were taken off the nutritional labels of each package.
The edited spectral
file was imported into NTAS(NIR Technology Australia Software) where the
spectra were examined. It can be seen that the chocolate samples absorbed very
strongly below 820nm. Normally spectra should be between 1 and 4 absorbance
units, ie, between 10 and 0.01% energy throughput. Spectra with absorbance
above 4 and especially 5, should not be used for calibration.
The Calibration
routine in NTAS uses Partial Least Squares(PLS) regression to derive a
calibration model for each component between the spectra and the component
levels. Figures 2 through 5, show the calibration plots for fat, protein,
carbohydrates and solids.

Figure 1: NIT Spectra
of Ice Cream Samples

Figure 2: Calibration
Plot – Fat

Figure 3: Calibration
Plot – Protein

Figure 4: Calibration
Plot – Carbohydrates

Figure 5: Calibration
Plot - Solids
Discussion:
The calibration plots
for fat and solids demonstrate excellent linearity and low errors. This data
indicates that the Series 3000 Food Analyser is capable of measuring fat and
solids in ice cream. The plots for protein and carbohydrates are not as strong
as for fat and solids. The range of protein values is too small to properly
develop a calibration. And the distribution of carbohydrate in these samples is
skewed to 21 to 23%. More samples between 23 and 27% would be required to
develop a reasonable calibration. Nonetheless the errors of calibration for all
four components are considered in line with other analytical techniques.