Application Note 46: Analysis of Fat and Moisture in Chocolate Blend Milk Powder
Introduction:
The NIT-38 Dairy Analyser provides the ability to develop
calibrations for a broad range of dairy related products, eg, milk, whey
powder, milk powder, cheese and yogurt. The instrument measures the NIR light
transmitted through a sample of liquid, powder or semi solid. At specific
frequencies, protein, fat and moisture absorb the light. Partial Least Squares
calibrations can be developed and then downloaded to the instrument for routine
analysis.
Description:
11 samples of Chocolate Blend Milk Powder were scanned
across the wavelength range, 720-1100nm. % scans were collected for each sample
in order to average out the distribution of the chocolate in the powder. The
spectra were loaded into NTAS(NIR Technology Australia Software) where PLS
calibrations were developed for both fat and moisture.
Results:
Figure 1 shows the NIR spectra of the Chocolate Blended Milk
Powders.

Figure 1. NIR spectra of Chocolate Blended Milk Powders

Figure 2. Plot of NIR Fat vs Ref Fat%

Figure 3. Plot NIR Moisture vs Ref Moisture %
Figures 2 and 3 show the plots of the Fat and Moisture
calibrations.
The calibration statistics are as follows;
Fat Moisture
# Sample Scans 55 55
# PC’s 12 15
Std Error of Cal (SEC) 0.09 0.09
R2 0.98 0.84
Comments:
The range of fat and moisture values for these samples is
too narrow. A superior calibration will be possible with samples covering at
least a +/- 2% range for each constituent. It is appreciated that production samples
rarely show such variation. As such, it is suggested that samples be blended to
give a broader range.
Conclusion:
Although the number of sample sis small for this study, the
data shows that the NIT-38 Dairy Analyser can make measurements of milk powders
with excellent agreement to the reference data. It would be expected that the
NIR technique could provide an accuracy of between 0.1% and 0.2% for both fat
and moisture in this type of sample.