NIR Technology Australia Suite 103, 56 Kitchener Pde
Bankstown,
NSW, 2200
Ph:
02 9708 5068
Fax:
02 9708 5537
Application Note
23: Measurement of Whole Milk using the
NIT-38 Dairy Analyser
Introduction:
The NIT-38 Dairy
Analyser provides a means of analyzing a broad range of dairy products for
components, such as, protein, fat and moisture. The NIT-38 is a NIR
transmission spectrophotometer based on diode array technology. The analyzer
scans the NIR spectrum from 720 to 1100nm. Within this region, C-H (fat), N-H
(protein) and O-H (water), absorb NIR energy. The analyzer measures the amount
of light absorbed by a sample as the light passes through it. The transmitted
light is separated into its frequency domain using a concave holographic
grating and a silicon photo diode array detector. Partial Least Squares (PLS)
regression techniques are used to correlate the NIR spectra with the
concentration of protein, fat and moisture in the samples. Calibration models
are then downloaded into the NIT-38 Analyser. Analysis of routine samples is as
easy as loading the sample cell, selecting the product type and following the
instructions on the screen. Analyses of up to 4 constituents are provided
within 60 seconds.
Description:
10 samples of
whole milk were scanned in duplicate on the NIT-38 Dairy Analyser using a
Liquid Cell with a 5mm pathlength. 3 spectra were collected for each sample
scan. The spectra were stored in the analyser’s memory and later transferred to
a PC. The spectra of several samples of milk are shown below.
Analyses for Fat,
Protein and Lactose were provided based on a Foss Milkoscan, were combined with
the spectral data in Microsoft Excel. The file was loaded into NTAS(NIR
Technology Australia Software) suite and a PLS calibration performed on each
constituent.
The same 10
samples were scanned and used as a prediction set.
Results:
The attached four
figures shows the prediction data for each constituent.
The following
table summarises the prediction set data.
Constituent Fat Protein Lactose
SEC 0.012 0.02 0.016
R2 0.99 0.96 0.98
SEP 0.027 0.032 0.04
Discussion:
It must be stated
that the prediction set is not a valid test of the accuracy of these
calibrations, since the set contains the same samples used for calibration.
Unfortunately there were insufficient samples provided to make a separate
prediction set.
The calibration
statistics for fat, protein and lactose show excellent correlation and low
errors. The statistics for the prediction set show higher errors. This is
considered to be caused by insufficient
samples to develop a robust calibration. It is recommended that between
50 and 100 samples of whole milk be scanned with as broad a constituent range as
possible. As well, samples covering wider regional selection as well as over
many months would be beneficial.
The data presented
above demonstrates that the NIT-38 Dairy Analyser is capable of measuring fat,
protein and lactose in milk.

Milk Spectra

Milk Fat
Prediction

Milk Protein
Prediction

Milk Lactose
Prediction